Reading a flow curve
A flow curve, steady-shear viscosity η plotted against shear rate γ̇ on log–log axes, is one of the quickest overall portraits of a liquid formulation. Before fitting a model, look at the shape.
The regions
- Zero-shear plateau (left). At low enough rates most polymer solutions and melts show a constant viscosity, η₀. For linear, flexible polymers above the entanglement molar mass, η₀ scales approximately as M3.4, which makes it a sensitive, if indirect, indicator of molar mass.
- Shear thinning (middle). Above a critical rate, chains or particle structures cannot relax as quickly as they are deformed, and viscosity decreases, often approximately as a power law. The onset rate is of the order of the inverse of the longest relaxation time.
- Second plateau (right). At very high rates the viscosity may level off again. This region is often beyond the range of a rotational rheometer, but it is relevant to processes such as spraying and high-speed coating.
Matching the process to the curve
Each process probes the curve at its own shear rate, so a viscosity value is meaningful only with its shear rate (and temperature). Typical orders of magnitude quoted in rheology handbooks are: sedimentation and levelling 10-3–10-1 s-1; pouring around 101 s-1; extrusion 102–103 s-1; coating and spraying 103–105 s-1. These are rough guides; the actual rate depends on geometry and speed. A dispersion that seems thick in the jar (low rate) and spreads easily under a doctor blade (high rate) is behaving as expected for a shear-thinning material.
Two traps
Yield-like behaviour. If viscosity keeps rising as the rate decreases, with no plateau in the measured window, the sample may show an apparent yield stress (gels, highly filled systems). A zero-shear viscosity extrapolated from such data is not reliable; a stress-controlled ramp or creep test is more informative.
Time dependence (thixotropy). If increasing and decreasing sweeps disagree, the structure changes with time under shear, and a single curve does not describe the material on its own. Apply a consistent pre-shear and rest period, and report the protocol with the data.
Two inks have the same viscosity at 1 s-1. Do they print the same way? Not necessarily: printing operates at much higher shear rates, where their flow curves may differ.
References
- C. W. Macosko, Rheology: Principles, Measurements, and Applications, Wiley-VCH, 1994.
- H. A. Barnes, A Handbook of Elementary Rheology, University of Wales Institute of Non-Newtonian Fluid Mechanics, 2000.